Крючков Фундаменталс оф Нуцлеар Материалс Пхысицал Протецтион 2011
.pdfR=240 Pueff (473fis/s ×g)ε 2 exp(-P /τ ) ´
´[1- exp(-G /τ )]∑P(ν ) |
ν (ν -1) |
(5.22) |
|
, |
|||
2 |
|||
ν |
|
where Р is the pulse count predelay time, G is the coincidence count time; τ is the neutron lifetime in the detector, ε is the neutron detection efficiency,
νis the number of neutron emitted by fission, and P(ν) is the probability of
νneutrons to be emitted by fission.
A schematic of a passive neutron coincidence counter for measurements of small-size samples is shown in Fig. 5.20.
Preamplifier
П
Lids
3Не-counter
П
Cavity Sample
Cadmium
Polyethylene
Fig. 5.20. Schematic of a passive neutron coincidence counter for measurements of small-size samples
The neutrons emitted by the sample are slowed down in the polyethylene and detected by 3Не-counters. The cavity for the samples is cadmium-shielded against slow neutrons, coming back from the polyethylene, to reduce the sample self-screening.
The counter operates in two modes: for thermal and fast neutrons. For fast neutron count, the sample cavity walls are coated with cadmium. Measurements in the fast-neutron mode fit better large-mass samples and thermal-neutron measurements are fit for small-mass samples. Thermalneutron measurements entail a smaller statistical error of small-size sample control. For large-size samples, a great cross-section value leads to the inner space being screened and the result distorted.
261
The counter is calibrated to interpret the analysis results. Calibration curves for the fastand thermal-neutron modes differ greatly. The calibration curve for the fast-neutron mode comprises two different segments: a segment with the influence of self-screening present (samples of the mass up to 500 g of 235U), and a further segment with multiplication where the 235U mass is rather large to compensate for the self-screening effect thanks to auxiliary fissions. Calibrating the interval of 150–900 g of 235U requires more than one standard. Each material type needs a special curve (Fig. 5.21).
.
time unit
events Coincidence count rate
500
400
300
200
100
0
0 |
20 |
40 |
60 |
80 |
100 |
Weight of 235U, g
Fig. 5.21. Coincidence count rate depending on the 235U mass for low-enriched U3O8 samples in the thermal neutron detection mode
Table 5.12 gives characteristics of an active well coincidence counter (AWCC).
There is a great variety of instruments based on neutron coincidence count, still all of them have standard electronic components.
Both neutron and gamma ray measurements involve a problem of measuring lengthy NM samples. Passive measurements of lengthy samples require ensuring similar probability conditions for detection of neutrons
262
emitted by all sample surface elements, while active measurements additionally require the same irradiation of all these elements by source neutrons. So every effort is made in designing neutron measuring systems to ensure a uniform sensitivity and a uniform field of external source neutrons in the cavity for the samples.
|
|
|
|
Table 5.12 |
|
Data of active well coincidence counter (AWCC) |
|||||
|
|
|
|
|
|
Characteristics |
Thermal mode |
|
Fast mode |
|
|
Mass of measured samples |
up to 100 g of 235U |
100–20000 g of 235U |
|
||
Coincidence count rate for low- |
11 count/(s×g of |
235 |
U) |
0.18 count/(s×g of |
|
enriched U3O8 sample |
|
235U) |
|
||
Absolute measurement error for |
0.3 g 235U |
|
|
18 g 235U |
|
large-size samples for 1000 s |
|
|
|
|
|
Calorimetry
Calorimetry is a passive nondestructive technique of NM (plutonium and tritium) control based on accurate temperature measurements. Generally, this is a more accurate technique, still it requires good temperature stability and control thereof, and is less fast and handy as compared to other NM nondestructive measurement methods.
Calorimeter is an instrument to measure the heat quantity emitted by an object.
Calorimetry offers an advantage that measurement results do not depend on the sample geometry, the matrix material or the NM distribution inside the sample. No standards identical to samples are required for calibrations. Calorimetric analyses have the accuracy comparable to that of chemical analyses .
Method description
All energy of α-decays transforms into heat. Each α-decay is accompanied by the energy yield of Q = (M 240 Pu − M 236 Pu − Mα )c2 =
Measurements of homogeneous burnt-up Pu samples give the accuracy of 0.1% as in chemical analyses and weighing. Measurements of waste containing Pu of a uniform isotopic composition gives a 1% accuracy.
263
5.25578 MeV. b-decay of 241Pu yields Qβ = 20.81 keV, and b-decay yields 3H Qβ = 18.59 keV.
Radioactive decays of 240Pu liberate P = l×N×Q of power, where N is
the number of 240Pu atoms, and l is the 240Pu decay constant. a-decay of 240Pu liberates 0.00707±0.00002 W/g of power. The total power generated by
all plutonium isotopes is Рeff(W/g)=ΣfiPi, where fi is the mass fraction of a single isotope.
Table 5.13 gives an example of the contribution made by some isotopes to Рeff. Рeff increases with the Pu burn-up increase. Burn-up fraction is characterized by the content of 240Pu.
|
|
Table 5.13 |
|
Contribution of selected isotopes to Рeff for one of the samples |
|||
|
|
|
|
|
Content, mass fraction |
Contribution to heat |
|
Isotope |
|
generation, % |
|
|
|
|
|
238Pu |
0.0006 |
11.0 |
|
239Pu |
0.8567 |
53.3 |
|
240Pu |
0.1211 |
27.7 |
|
241Pu |
0.0194 |
2.1 |
|
242Pu |
0.0022 |
0.0 |
|
241Am |
0.0016 |
5.9 |
|
By measuring the heat generated by a plutonium sample and knowing its isotopic composition, one can find the content of Pu.
The sample-generated heat is recorded by a heat sensor out of a sensitive wire laid in circles around the sample cavity. A double calorimetric bridge with two identical thermostats is shown in Fig. 5.22.
Measurements are done using a potentiometer or a digital voltmeter incorporated in an electric circuit called Wheatstone bridge (Fig. 5.23). The measured voltage is proportional to the difference between the temperature in the sample space and the temperature of the comparison probe, the latter
being in an air or water “bath” with a constant tem perature (maintained within ±0.001 °C).
264
|
Sample thermostat |
Comparison sensor |
|
thermostat |
|
|
|
|
Resistance bridge |
|
|
nickel |
|
|
winding |
|
|
Heat chamber |
|
|
wall |
|
Sample |
Air |
|
cavity |
|
|
|
gap |
|
|
Plastic |
|
Heat-insulating |
end |
|
material |
Fig. 5.22. Schematic of a double calorimetric bridge with two identical thermostats
V
Reference arm
Current
source
Reference arm
Reference arm |
Test arm
Fig. 5.23. Wheatstone bridge for heat flux measurements
If the temperature in both thermostats is identical, that is there is no sample, the Wheatstone bridge is balanced. When the sample is placed in the thermostat, the temperature changes and the bridge turns unbalanced.
The voltmeter indication is taken some time after the sample is placed in the thermostat. The time needed to have equilibrium reached depends on the sample dimensions and amounts to hours (Fig. 5.24). Preheating of the
265
thermostat chamber for the sample helps reach the equilibrium in several times as fast.
To determine the power generated by the sample in the thermostat from the measured potential difference, a sensitivity diagram is normally used.
Equilibrium value |
Bridge voltage, V |
Time, h |
Fig. 5.24. Time-dependent voltmeter indications
To this end, a curve for the thermostat sensitivity to the samplegenerated power is plotted. Fig. 5.25 shows a calorimeter sensitivity diagram option.
V/W |
|
|
|
S, |
|
|
W, W |
|
|
|
|
10 |
20 |
30 |
40 |
Fig. 5.25. Calorimeter sensitivity diagram
For calibration, the calorimeter is switched on and the potential difference at the bridge ends (ВР0) is measured without a sample or any other heat source. Then a plutonium standard is placed in the sample
266
chamber, the equilibrium value for the potential difference ВРs is measured and the calorimeter sensitivity is calculated by the formula:
S = (BPS – BP 0)/WS , |
(5.23) |
where WS is the standard-emitted power (not known).
Normally, the quantity S depends slightly on the heat source power. For example, a 1.6% sensitivity decrease was recorded in measurements of samples with the power of 0.1 to 10 W.
As the diagram in Fig. 5.25 shows, sensitivity in the given case is not constant and depends on the sample-generated power.
By and large, the best of the calorimetric analyses have the following
errors: power measurement error <0.1%, effective specific energy release determination error <0.2%.
Typical calorimeter parameters: diameter – 120 mm, height – 275 mm, range – 0–6 W.
Referenced standards or electrical standards (probes) are used to calibrate calorimeters:
1)heat generation standards – 238Pu samples. These feature:
∙small dimensions (enable determination of errors depending on the heat distribution over the chamber);
∙qualification accuracy – 0.02%;
∙decay may be accurately taken into account;
2) electrical heat generation standards. These feature:
∙absence of radioactive radiation;
∙no need for decay to be taken into account;
∙electronics may not depend on the calorimetric system;
∙electronics requires calibration.
The following is taken into account in calorimeter design.
∙Sample size (specifies the sample chamber size). A close samplecalorimeter contact makes it possible to minimize the analysis time. The chamber diameter in existing calorimeters varies form 1 to 30 cm.
∙Sample heat power. High-power samples need low-sensitivity calorimeters with a low heat resistance, and small-power samples require high-sensitivity and highly heat-resistant calorimeters.
∙Graduation techniques. The calorimeter design depends on what heat source is used for graduation (a radioisotope or an electrical one).
∙Capacity. Selection of the calorimeter type depends on the analysis time required.
267
∙Accuracy. When the calorimeter type and its operating mode are selected, the analysis accuracy is planned given the time to be spent and other conditions.
∙Application environment. Selection of the calorimeter design depends on the environment and the available workroom area to install it.
5.3. Destructive analyses
Normally, a destructive analysis (DA) includes sample taking, sample chemical treatment and measurement stages. Test material may be a single item or a bulk material. Destructive analysis normally offers a higher accuracy of results than nondestructive measurements. However, DA takes more time to perform and is more costly than nondestructive tests.
Destructive analyses are used to:
∙check nondestructive measurement data;
∙perform critical test measurements;
∙qualify standards.
Sample taking
Destructive analyses have a small portion of controlled material examined and require a representative sample to be taken for analysis. The sample composition should be strictly representative of the average material composition with the sample mass (volume) accurately determined.
If the sample is soluble, it needs to be, generally, dissolved. Sometimes, uranium or plutonium requires separation from interfering elements.
Sample should be taken given a potential heterogeneity of NM, which can be of three types:
∙heterogeneity of material within the container;
∙differences among containers;
∙differences among groups of containers.
Some dispersed and powdered materials, such as ash and roasted scrap, are hard to mix. A number of factors exist that require homogenization of dry powders and dispersed compositions. These are, for example:
∙variations of the composition depending on the particle size and density;
∙differences in the particle forms;
∙cohesion and conglomeration of particles.
268